6253 DC Voltage Current Source/Monitor Operation Manual
3.2.1 MENU Key (Parameter Setting)
3-8
6) Measure Sets the measurement-related parameters.
Auto Zero Switches the measurement Auto Zero function ON or OFF.
ON: Corrects measurement zero-point drifts approximately
every 10 seconds. The AZ indicator turns ON.
OFF: Turns OFF the Auto Zero function. The AZ indicator
turns OFF.
Integ Time Sets the measurement integration time.
Select the integration time from
5 s, 10 s, 100 s, 500 s, 1 ms, 10 ms, 1 PLC, 2 PLC, 200 ms,
and Variable.
Variable IT Sets the variable integration time between 100 s and 1000 ms.
Measure SW Switches measurement ON or OFF.
ON: Executes measurement.
OFF: Not execute measurement.
Disp Digit Selects the number of measurement display digits.
Spaces are displayed as blank digits but do not affect any
measurement data.
6 digits: Displays measurement data with 6
1
/
2
digits.
5 digits: Displays measurement data with 5
1
/
2
digits.
4 digits: Displays measurement data with 4
1
/
2
digits.
3 digits: Displays measurement data with 3
1
/
2
digits.
Disp Unit Selects the output format style for measurement data and
comparator upper and lower limit values.
Prefix: Displays measurement data by using a unit symbol.
Exponent: Displays measurement data in an exponential form.
Mfunc Link Links the measurement function to the source function.
ON: The source function is linked to the measurement
function. (VSIM/ISVM)
OFF: The measurement function is unaffected by the source
function.
Sample/Trigger Performs sampling the specified number of times by one trigger
when the sampling mode is HOLD.
The sampling count is settable between 1 to 20,000. (Not
available in the sweep source mode.)