6253 DC Voltage Current Source/Monitor Operation Manual
3.2.1 MENU Key (Parameter Setting)
3-12
TRIG IN Sets the external trigger input.
Disable: Disables the external trigger input.
Enable: Enables the external trigger input.
Sig Width Selects the output pulse width, 10 s or 100 s.
11) Parameter Loads or saves setting parameters.
Param Load Loads saved parameters save in the non-volatile memory. Not
available in the Operate status.
Load0 to 3: Loads the setting parameters from areas Save0to
3.
LoadDefault: Loads the factory default settings as setting
parameters.
Param Save Saves the setting parameters.
Save0 to 3: Saves the setting parameters into areas Save0 to
3 of the non-volatile memory.
SetDefaultALL:
Sets the factory default settings in all the areas
from 0 to 3.
12) Interface Selects or sets the interface.
I/F BUS Selects the interface.
USB: Selects the USB interface.
GPIB: Selects the GPIB interface.
LAN: Selects the LAN interface.
RS232: Selects the RS-232 interface.
Header Switches the header ON or OFF.
Output Monitor Switches the source value monitor ON or OFF.
GPIB Address Sets the GPIB address from 0 to 30.
USB ID Sets the USB ID from 1 to 127.
RS232 Config Sets the RS-232 configuration.
Baud Rate Selects the baud rate.
19200: 19,200 (bps)
9600: 9,600 (bps)
4800: 4,800 (bps)
2400: 2,400 (bps)
1200: 1,200 (bps)
600: 600 (bps)
300: 300 (bps)