WARNING:
If a grinding noise is heard during movement of the X-ray tube
head unit or radiographic wall stand, the steel cables inside the
ceiling suspension or wall stand could be broken. Do not operate
the unit any further and try to avoid hard vibration or knocks of
any kind. Please contact service.
Note: If movement in any direction blocks, do not apply force to
overcome the blocking. Contact your local service organization.
Note: To avoid shock and damage, move the tube head unit with
normal speed and slow down when reaching the mechanical end
stops.
Note: Rotation may be limited by cables. Avoid strain on the cables
during rotation.