AGFA DR 600 - No table movement

AGFA DR 600
315 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
No table movement
Details The table is not moving up or down when pressing the
foot pedals with double click. No error is shown.
Cause One of the foot pedals was pressed longer than 90 sec-
onds.
Brief Solution
1. Press the Power OFF button on the X-ray generator
control box to switch off the generator.
2. Switch off the electrical room switch.
3. Wait for 30 seconds.
4. Switch on the electrical room switch.
5. Press the Power ON button on the X-ray generator
control box to switch on the system.
DR 600 | Problem solving | 259
3261K EN 20210708 0808