Generator Technical Data (EMD)
Manufactur-
er
EMD Technologies
400 du Parc,
Saint-Eustache, (Quebec) Canada, J7R 0A1
Supported
Models
EPS 50 EPS 65 EPS 80
Power rating
for 380 VAC
105 A / 68 kVA 135 A / 85 kVA 165 A / 105 kVA
Power rating
for 400 VAC
102 A / 68 kVA 124 A / 85 kVA 150 A / 102 kVA
Power rating
for 480 VAC
85 A / 70 kVA 105 A / 88 kVA 125 A / 105 kVA
Max. Power
Output
50 kW 60 kW 80 kW
Power Out-
put (at 0.1s)
400 mA: 125 kVp
500 mA: 100 kVp
630 mA: 79 kVp
500 mA: 130 kVp
630 mA: 103 kVp
800 mA: 81 kVp
1000 mA: 65 kVp
630 mA: 127 kVp
800 mA: 100 kVp
1000 mA: 80 kVp
kV-Range 40-150 kV
mAs-Range 0.4-1000 mAs
mA-Range 10-630 mA 10-800 mA 10-1000 mA
ms-Range 1-10000 ms
Power line
380 V
380 VAC ± 10% 50/60 Hz (3 Phases)
Power line
400 V
400V VAC ± 10% 50/60 Hz (3 Phases)
Power line
480 V
480V VAC ± 10% 50/60 Hz (3 Phases)
The power setting is selected during installation and printed
on the type label.
Dimensions 80.6 cm x 47.2 cm x 77.6 cm (HxWxD)
Weight 79-90 kg depending on the configuration
Duty cycle The Generator duty cycle is continuous, but limits should be
set during installation depending on the capacity of the X-ray
tube.
284 | DR 600 | Product Information
3261K EN 20210708 0808