Stopping the System
To stop the system:
1. Stop the NX workstation.
NX can be stopped in two ways, either by logging out of Windows or
without logging out of Windows.
For detailed information on stopping NX, refer to the NX User Manual,
document 4420.
Note: Stopping the NX workstation does not stop the DR
Detector. If the power of the DR Detector remains on, no
warming-up time will be needed after starting the NX
workstation.
2. Press the Power OFF button on the X-ray generator control box to switch
off the generator.
3. In a configuration with a wireless DR Detector, power off the DR Detector:
turn off the DR Detector.
remove the battery pack.
4. Switch off the DR Generator Sync.
Note: If the DR Detector is powered down, a warming-up may be
required on the next start-up.
Related Links
Moving the system to parking position on page 165
Power off behaviour on page 43
DR 600 | Basic Workflow | 121
3261K EN 20210708 0808