Restoring connection between generator and NX after
generator failure
Details An error on the generator occurred. NX lost connection
to the generator.
An error message that no connection with the genera-
tor can be established is displayed on the Software
Console.
Cause After a shutdown of the generator, the communication
between the X-ray generator and the NX workstation is
broken.
Brief Solution To set up the communication between the X-ray gener-
ator and the NX workstation:
1. Switch off the X-ray generator at X-ray generator
console.
2. After some seconds, switch the X-ray generator
back on.
3. Select an empty thumbnail in the Image Overview
pane of the Examination window.
4. The error message disappears. This may take some
time.
If an error is indicated on the X-ray generator by a sig-
nal, repeat step 1 to 3.
During startup of the NX application and the Software
Console, the communication to the generator is set up
and the self-test of the generator is triggered.
256 | DR 600 | Problem solving
3261K EN 20210708 0808