EasyManua.ls Logo

AGFA DR 600 - Dose Area Product Meter (IBA DAP) Technical Data

AGFA DR 600
315 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dose Area Product Meter (IBA DAP) Technical Data
Manufacturer IBA Dosimetry GmbH
Bahnhofstrasse 5
DE-90592 Schwarzenbruck
Supported Type
120-131 HS/RS485
Dose area product range (0.1...99999999.99) cGy x cm
2
Minimum readout value 0.01 cGy x cm
2
Active area
14.0 cm x 14.0 cm
Dimensions 17.9 cm x 16.6 cm x 1.7 cm (WxDxH)
Weight approx. 220 g
Correction factors for using the DAP meter on high altitude
Environmental conditions Correction factor
75 kPa (ca. 2500 m)
0° Celcius
1.26
75 kPa (ca. 2500 m)
20° Celcius
1.35
70 kPa (ca. 3000 m)
0° Celcius
1.35
70 kPa (ca. 3000 m)
20° Celcius
1.45
DR 600 | Product Information | 295
3261K EN 20210708 0808

Table of Contents

Related product manuals