Generator Technical Data (Spellman)
Manufacturer
Spellman High Voltage Electronics GmbH
Josef-Baumann-Strasse 23
D-44805 Bochum, Germany
Supported
Models
EDITOR HFe 501 EDITOR HFe 601 EDITOR HFe 801
Power rating
for 400 VAC
113 A / 76 kVA 144 A / 96 kVA 180A / 120 kVA
Power rating
for 480 VAC
97 A / 76 kVA 124 A / 96 kVA 154 A / 120 kVA
Max. Power 50 kW 65 kW 80 kW
Power Output
(at 0.1s)
625mA: 80kVp
500mA: 100kVp
400mA: 125kVp
330mA: 150kVp
800mA: 80kVp
650mA: 100kVp
520mA: 125kVp
430mA: 150kVp
800mA: 80kVp
800mA: 100kVp
640mA: 125kVp
530mA: 150kVp
kV-Range 40-150 kV
mAs-Range 0.5-600 mAs
mAs-Range
(tomosynthe-
sis exposures)
0.1-20 mAs
Maximum dose (air kerma) is reached at SID 100 cm, 150
kV, 16 mAs
mA-Range 10-650 mA 10-800 mA
ms-Range 1-6300 ms
Power line
400 V Y-
source
400V 3N~ PE (Y) 50/60 Hz
Power line
400/480 V
Delta-source
400/480V 3~PE (delta without N) 50/60Hz
The power setting is selected during installation and printed
on the type label.
Dimensions 97 cm x 55 cm x 63 cm (HxWxD)
Weight 129 kg
Duty cycle The Generator duty cycle is continuous, but limits should be
set during installation depending on the capacity of the X-
ray tube.
DR 600 | Product Information | 283