EasyManua.ls Logo

AGFA DR 600 - Dose Area Product Meter (Vacutec DAP) Technical Data

AGFA DR 600
315 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Dose Area Product Meter (VacuTec DAP) Technical Data
Manufacturer VacuTec Messtechnik GmbH
Dornblüthstrasse 13
D-01277 Dresden, Germany
Supported Type
VacuDAP 2004
Dose area product range (1.0...9999999.9) cGy x cm
2
Minimum readout value 0.1 cGy x cm
2
Active area
14.7 cm x 14.7 cm
Dimensions 18.2 cm x 17.7 cm x 1.8 cm (WxDxH)
Weight 270 g
Correction factors for using the DAP meter on high altitude
Environmental conditions Correction factor
75 kPa (ca. 2500 m)
0° Celcius
1.26
75 kPa (ca. 2500 m)
20° Celcius
1.31
70 kPa (ca. 3000 m)
0° Celcius
1.35
70 kPa (ca. 3000 m)
20° Celcius
1.40