Dose Area Product Meter (DAP)
An integrated DAP meter (Dose Area Product Meter) in the automatic
collimator is availabe as an option.
The DAP meter reads the radiation as Dose Area Product in [cGy x cm
2
].
The measured radiation value is transferred to the Software Console
automatically and displayed after each exposure. No value is displayed if the
measured radiation value is below the minimum readout value of the DAP
meter.
The DAP meter cannot be removed from the collimator.
The DAP meter is calibrated during production to be used up to 2000 m
altitude. Using the DAP meter on higher altitudes requires the application of a
correction factor.