Equipment Classification
Per EN/IEC 60601-1, EN/IEC 60601-2-54, this device is classified as
following:
Table 43: Equipment classification
Class I equipment
Equipment in which protection against electric
shock does not rely on basic insulation only, but in-
cludes a fixed connection to mains power with pro-
tective earth conductor.
Type B equipment A Type B applied part is one that provides a partic-
ular degree of protection against electric shock
particularly regarding allowable leakage current
and reliability of the protective earth protection.
Water ingress IP10
This device does not have protection against in-
gress of water.
Cleaning See section on cleaning and disinfecting.
Disinfection See section on cleaning and disinfecting.
Flammable anesthetics This device is not suitable for use in the presence of
a flammable anesthetic mixture with air, or in
presence of a flammable anesthetic mixture with
oxygen or nitrous oxide.
Operation Continuous operation.
Related Links
Cleaning and Disinfecting on page 74
DR 600 | Product Information | 271