No movement when using remote control
Details No movement of the system can be activated using the
remote control. No error is shown.
Cause Communication between the remote control and the
system.
Brief Solution
1. Press the Power OFF button on the X-ray generator
control box to switch off the generator.
2. Switch off the electrical room switch.
3. Wait for 30 seconds.
4. Switch on the electrical room switch.
5. Press the Power ON button on the X-ray generator
control box to switch on the system.
260 | DR 600 | Problem solving