Changing the grid configured with an earthquake kit
If the system is configured with an earthquake kit, the grid is locked in the
bucky by a latch mounted on the grid. Release the latch to pull out the grid
from the bucky.
1. Release the latch.
Figure 115: Latch mounted on the grid
2. Pull out the grid from the bucky using the handle.
A magnet inside the bucky provides some resistance for pulling out the the
grid, even when the latch is released.