Active pixel matrix 3056(H) x 3056(V)
Fill factor 100 %
Detector type Amorphous Silicon
Active area size 42,7 cm (H) x 42,7 cm (V)
Reliability
Estimated product life (if regularly serviced
and maintained according to Agfa instruc-
tions)
100 000 RAD
Fixed DR Detector Technical Data
Manufacturer
Manufacturer DR Detector THALES AVS FRANCE SAS
460 Rue du Pommarin – BP122
38430 MOIRANS
France
Supported models
Pixium RAD 4343 C (Pixium 4343RC)
Pixium RAD 4343 C-E
CsI conversion screen
Pixium RAD 4343 G (Pixium 4343RG)
Pixium RAD 4343 G-E
GOS conversion screen
Electrical Connection
Operating voltage +24V 3.5A DC
Warming-up time
5 minutes
Throughput
Maximum number of image acquisitions 150 acquisitions per hour
Reliability
Estimated product life (if regularly serviced
and maintained according to Agfa instruc-
tions)
100 Gy
298 | DR 600 | Product Information
3261K EN 20210708 0808

Table of Contents