CAUTION:
Excessive ambient temperature may impact performance of DR
Detectors and cause permanent damage to the equipment. Refer
to the related user manual for environmental conditions for the
DR detector. If ambient temperature and humidity is outside the
specified range, do not operate the system or use air
conditioning. Frost due to low temperatures can damage
internal circuits. Warranty will be void if it is obvious that
operating conditions are not met.
CAUTION:
To avoid images being lost due to a power failure, the
workstation and the Digitizer have to be connected to
uninterruptable power supply (UPS) or an institutional standby
generator. In case of a power failure, the UPS will allow to
finalize exposed images that are being scanned.
CAUTION:
Install the NX workstation and CR digitizer at a minimum (safe)
distance of 2 m from the X-Ray System components or provide a
wall or window to separate both systems.