6 Method Parameters
Set the 2D-LC Method parameters
2D-LC User Guide 138
Define Parameters for Time-Based Comprehensive
Parameters for Time-Based Comprehensive
Figure 75 Comprehensive preview with a modulation time 0.3 min (=20 cycles)
1 Enter an absolute time range where the system creates equidistant cuts.
Comprehensive run starts at the given time.
The modulation time determines the cut size.
2 Enter the stop time of the comprehensive measurement.
Comprehensive Range Stop at, e.g., 10.0 min.
The stop time should coincide with that of the 2D pump.
NOTE
In comprehensive the function of the flush gradient is not available.
NOTE
If the sampling table is empty, no 2D-LC operation will be executed at all.