9 Data Analysis
2D-LC Data Analysis/Data Evaluation for MassHunter
2D-LC User Guide 234
MassHunter Qualitative Analysis Software
The Masshunter data analysis software generally works with
2
D data in the same
way as you are used to. The task to find an identify compounds or setup and run
qualitative analysis methods can also be performed on 2D-LC data. However, to
make it easier to get started with
2
D data, we have listed some different
workflows as examples.
The 2D-LC instrument in this case was equipped with 2 UV detectors (
1
D and
2
D),
2D-LC valve with MHC and a Q-TOF detector in the second dimension. Default
method is loaded.
Workflow
1
D UV Data Extraction – Alternative 1
1 Open Data File 2D-LC file.d, in this case Test_Comp r001.d.
The string of 2D TIC-chromatograms appears.