Agilent 4150 TapeStation System Manual 129
Maintenance
9
Exchange the Piercing Needle
6 Close Lid. Click Removed.
7 Wait approximately two minutes until the automated pierce needle
calibration is complete. It is crucial not to turn off the instrument or
terminate the software during the needle exchange; this could impact
performance of all future analyses.