144 Agilent 4150 TapeStation System Manual
9
Maintenance
Electrode Cartridge
21 Reinsert the TapeStation Test Tape.
22 Close the lid.
23 Click Proceed to start a short verification.
24 Click Proceed to start System Verification tests. If the Electrode Cartridge
Change was not successful, please refer to “Troubleshooting” on page 165.