Agilent 4150 TapeStation System Manual 145
Maintenance
9
Electrode Cartridge
25 Execute System Verification tests and review the results to verify
instrument functionality.
Please be aware of software notifications that are related to the Electrode
Cartridge:
“Electrode Cartridge Change Required” on page 174
“Used Electrode Cartridge Inserted” on page 175
“Electrode Cartridge Change has Failed” on page 176
“Electrode Cartridge not Detected” on page 177