Agilent 4150 TapeStation System Manual 95
Good Measurement Practices
7
Mixing Recommendations
Figure 22 The electropherogram (A) and the gel image (B) of the ScreenTape reagent
mixing tests.
The green trace shows the recommended protocol of vortex mixing using the
IKA vortexer and adaptor at 2000 rpm for 1 minute followed by brief
centrifugation. Blue shows results for pipette mixing only. Red shows the
effect of no mixing. Images were taken from the TapeStation Analysis
software.