EasyManua.ls Logo

Agilent Technologies 4150 TapeStation system - Page 95

Agilent Technologies 4150 TapeStation system
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 4150 TapeStation System Manual 95
Good Measurement Practices
7
Mixing Recommendations
Figure 22 The electropherogram (A) and the gel image (B) of the ScreenTape reagent
mixing tests.
The green trace shows the recommended protocol of vortex mixing using the
IKA vortexer and adaptor at 2000 rpm for 1 minute followed by brief
centrifugation. Blue shows results for pipette mixing only. Red shows the
effect of no mixing. Images were taken from the TapeStation Analysis
software.

Table of Contents

Related product manuals