Chapter 9 689
If You Have a Problem
Error Messages
Eliminating Error Messages
It might be possible to eliminate some error messages by running the
LO and IF realignment procedure below, or by running the procedures
described in "Hardware Problems". If an error message remains
displayed, the following actions are suggested:
Error Action
100 to 199 Programming error detected. Refer to chapters 3 and 7
in the user's guide for information on programming the
spectrum analyzer.
200 to 799 The instrument is in need of service. Refer to this
chapter for information about calling Agilent
Technologies.
800 to 899 Option module failures detected. Refer to the
appropriate option-module for an explanation of these
errors.
900 to 999 User generated errors. Information entered incorrectly
or an attempt to use the spectrum analyzer
inappropriately.
Realigning the LO and IF
The LO and IF realignment procedure may eliminate error conditions.
1. Press
SAVE, then SAVE STATE.
2. Store the current state in a convenient STATE register.
3. Press
PRESET, CAL, then REALIGN LO &IF. Wait for the sequence to
finish.
4. Press RECALL, then RECALL STATE.
5. Recall the previously stored STATE.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8563E and is the answer not in the manual?

Agilent Technologies 8563E Specifications

General IconGeneral
BrandAgilent Technologies
Model8563E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals