690 Chapter9
If You Have a Problem
Error Messages
Error Code Listing
Error codes and their associated messages are listed in numeric order
below. Error codes 100 to 199 relate to incorrect spectrum
analyzer programming
ERR 100 NO PWRON Power-on state is invalid; default state is
loaded.
ERR 101 NO STATE State to be RECALLed not valid or not SAVEd.
ERR 102 # ARGMTS Command does not have enough arguments.
ERR 103 # ARGMTS Command does not have enough arguments.
ERR 104 # ARGMTS Command does not have enough arguments.
ERR 105 # ARGMTS Command does not have enough arguments.
ERR 106 ABORTED! Current operation is aborted; GPIB parser
reset.
ERR 107 HELLO ?? No GPIB listener is present.
ERR 108 TIME OUT Analyzer timed out when acting as controller.
ERR 109 CtrlFail Analyzer unable to take control of the bus.
ERR 110 NOT CTRL Analyzer is not system controller.
ERR 111 # ARGMTS Command does not have enough arguments.
ERR 112 ??CMD?? Unrecognized command.
ERR 113 FREQ NO! Command cannot have frequency units.
ERR 114 TIME NO! Command cannot have time units.
ERR 115 AMPL NO! Command cannot have amplitude units.
ERR 116 ?UNITS?? Unrecognizable units.
ERR 117 NOP NUM Command cannot have numeric units.
ERR 118 NOP EP Enable parameter cannot be used.
ERR 119 NOP UPDN UP/DN are not valid arguments for this
command.
ERR 120 NOP ONOF ON/OFF are not valid arguments for this
command.
ERR 121 NOP ARG AUTO/MAN are not valid arguments for this
command.
ERR 122 NOP TRC Trace registers are not valid for this command.
ERR 123 NOP ABLK A-block format not valid here.