Triggering the Oscilloscope 4
6000 Series Oscilloscope User’s Guide 129
Trigger modes: Normal and Auto
An introduction to Normal and Auto trigger modes is given on
page 78.
1 Press the Mode/Coupling key.
2 Press the Mode softkey, then select Normal or Auto.
• Normal mode displays a waveform when the trigger
conditions are met, otherwise the oscilloscope does not
trigger and the display is not updated.
• Auto mode is the same as Normal mode, except it forces the
oscilloscope to trigger if the trigger conditions are not met.
Auto mode
Use the auto trigger modes for signals other than
low-repetitive-rate signals and for unknown signal levels. To
display a DC signal, you must use auto trigger mode since there
are no edges on which to trigger.
When you select Run, the oscilloscope operates by first filling
the pre-trigger buffer. It starts searching for a trigger after the
pre-trigger buffer is filled, and continues to flow data through
this buffer while it searches for the trigger. While searching for
the trigger, the oscilloscope overflows the pre-trigger buffer; the
first data put into the buffer is the first pushed out (FIFO).
When a trigger is found, the pre-trigger buffer will contain the
events that occurred just before the trigger. If no trigger is
found, the oscilloscope generates a trigger and displays the data
as though a trigger had occurred. In this case, the background of
the Auto indicator at the top of the display will flash, indicating
that the oscilloscope is forcing triggers.
When you press the Single key, the oscilloscope will fill
pre-trigger buffer memory, and continue flowing data through
the pre-trigger buffer until the auto trigger overrides the
searching and forces a trigger. At the end of the trace, the
oscilloscope will stop and display the results.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSO6012A and is the answer not in the manual?

Agilent Technologies MSO6012A Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSO6012A
CategoryTest Equipment
LanguageEnglish

Related product manuals