Triggering the Oscilloscope 4
6000 Series Oscilloscope User’s Guide 151
< qualifier time set softkey
• When the less than (<) qualifier is selected, the Entry knob
sets the oscilloscope to trigger on a pattern duration less
than the time value displayed on the softkey.
• When the within time range (><) qualifier is selected, the
Entry knob sets the upper time range value.
• When the out of time range (<>) qualifier is selected, the
Entry knob sets the lower time range value.
> qualifier time set softkey
• When the greater than (>) qualifier is selected, the Entry
knob sets the oscilloscope to trigger on a pattern duration
greater than the time value displayed on the softkey.
• When the within time range (><) qualifier is selected, the
Entry knob sets the lower time range value.
• When the out of time range (<>) qualifier is selected, the
Entry knob sets the upper time range value.
• When Timeout qualifier is selected, the Entry knob sets the
timeout value.
When the duration trigger occurs
The timer starts on the last edge that makes the pattern (logical AND) true. The
trigger will occur on the first edge that makes the pattern false if the time
qualifier criteria of the pattern has been met, except in Timeout mode. In
Timeout, the trigger occurs when the timeout value is reached while the pattern
is true.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSO6012A and is the answer not in the manual?

Agilent Technologies MSO6012A Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSO6012A
CategoryTest Equipment
LanguageEnglish

Related product manuals