360 6000 Series Oscilloscope User’s Guide
9 Characteristics and Specifications
Storage
Save/recall (non-volatile) 10 setups and traces can be saved and recalled
internally. Secure environment mode (-SEC)
ensures setups and traces are stored to volatile
memory.
Storage type and format USB 1.1 drive on front (/drive0) and rear (/drive5)
panels
Image formats: BMP (8 bit), BMP (24 bit) and
PNG (24 bit)
Data formats: X and Y (time/voltage) values in
CSV, ASCII XY and binary format
Trace/setup formats: Recalled
I/O
Standard ports USB 2.0 high speed, 10/100-BaseT LAN,
IEEE488.2 GPIB, XGA video output
Max transfer rate IEEE488.2 GPIB: 500 kbytes/sec
USB (USBTMC-USB488): 3.5 Mbytes/sec
100 Mbps LAN (TCP/IP): 1 Mbytes/sec
General characteristics
Rack mounting Supplied with all necessary hardware (except
tools) for installation into a standard EIA 19-inch
rack
Physical size 43.5 cm W x 27 cm D x 4.2 cm H (without
brackets)
Weight Net: 2.45 kg (5.4 lbs.)
Shipping: 6.2 kg (13.6 lbs.)
Probe comp output Frequency ~1.2 kHz
Amplitude ~2.5 V
Trigger out
When Triggers is selected 0 to 5 V into high impedance
(delay ~17 ns) 0 to 2.5 V into 50 Ω
When Source Frequency or 0 to 580 mV into high impedance
Source Frequency/8 is selected 0 to 290 mV into 50 Ω
Max frequency output 350 MHz (in source frequency mode when
terminated in 50 Ω)
125 MHz (in source frequency/8 mode when
terminated in 50 Ω)
10 MHz ref in/out TTL out, 180 mV to 1 V amplitude within 0 to 2 V
offset