254 6000 Series Oscilloscope User’s Guide
6 Displaying Data
Antialiasing
At slower sweep speeds, the sample rate is reduced and a
proprietary display algorithm is used to minimize the likelihood
of aliasing.
By default, Antialiasing is enabled. You should leave
Antialiasing enabled unless there is a specific reason to switch
it off.
If you need to switch Antialiasing off, press
Utilities&Options&Preferences and press the Antialiasing softkey to
switch the feature off. The displayed waveforms will be more
susceptible to aliasing.
Using the XGA video output
A standard XGA video output connector is provided on the rear
panel. You can connect a monitor to provide a larger display, or
to provide a display at a viewing position away from the
oscilloscope.
The oscilloscope’s built-in display will remain on even when an
external display is connected.