Front-Panel Controls 2
6000 Series Oscilloscope User’s Guide 105
probes do not need to be calibrated, but InfiniiMax probes can
be calibrated. For more information about calibrating probes
see page 53.
Results:
User Cal date:
Change in temperature since last User Cal:
Failure:
Comments:
Probe Cal Status:
Self Test
Pressing Utility&Service&Start Self Test performs a series of
internal procedures to verify that the oscilloscope is operating
properly.
It is recommended you run Self Test:
• after experiencing abnormal operation.
• for additional information to better describe an oscilloscope
failure.
• to verify proper operation after the oscilloscope has been
repaired.
Successfully passing Self Test does not guarantee 100% of the
oscilloscope's functionality. Self Test is designed to provide an
80% confidence level that the oscilloscope is operating properly.
About Oscilloscope
Pressing Utility&Service&About Oscilloscope displays information
about your oscilloscope model number, serial number, software
version, boot version, graphics version, and installed licenses.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSO6012A and is the answer not in the manual?

Agilent Technologies MSO6012A Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSO6012A
CategoryTest Equipment
LanguageEnglish

Related product manuals