Characteristics and Specifications 9
6000 Series Oscilloscope User’s Guide 359
Measurement features
Automatic measurements Measurements are continuously updated.
Cursors track last selected measurement.
Voltage (scope channels only) Peak-to-peak, maximum, minimum, average,
amplitude, top, base, overshoot, preshoot, RMS,
standard deviation (AC RMS)
Time Frequency, period, + width, -width and duty cycle
on any channels
Rise time, fall time, X at max Y (time at max volts),
X at min Y (time at min volts), delay, and phase on
scope channels only
Counter Built-in 5-digit frequency counter on any scope
channel. Counts up to the scope’s bandwidth (1
GHz max). The counter resolution can be
increased to 8 digits with an external 10 MHz
reference.
Threshold definition Variable by percent and absolute value; 10%,
50%, 90% default for time measurements
Cursors Manually or automatically placed readout of
horizontal (X, DX, 1/DX) and vertical (Y, DY)
Additionally logic or scope channels can be
displayed as binary or hex values
Waveform math One function of 1-2, 1x2, FFT, differentiate,
integrate.
Source of FFT, differentiate, integrate: scope
channels 1 or 2, 1-2, 1+2, 1x2
FFT
Points Fixed at 1000 points
Source of FFT Scope channels 1, 2, 3 or 4, 1+2, 1-2, 1x2
Window Rectangular, flattop, Hanning
Noise floor -50 to -90 dB depending on averaging
Amplitude Display in dBV, dBm at 50 Ω
Frequency resolution 0.05/(time per div)
Maximum frequency 50/(time per div)