280 6000 Series Oscilloscope User’s Guide
6 Displaying Data
CAN Totalizer
The CAN totalizer provides a direct measure of bus quality and
efficiency. The CAN totalizer measures total CAN frames,
flagged error frames, overload frames, and bus utilization.
The totalizer is always running (counting frames and
calculating percentages) and is displayed whenever CAN
decode is displayed. The totalizer counts even when the
oscilloscope is stopped (not acquiring data). Pressing the
Run/Stop key does not affect the totalizer. When an overflow
condition occurs, the counter displays OVERFLOW. The counters
can be reset to zero by pressing the Reset CAN Counters softkey.
Types of Frames
• Active error frames are CAN frames in which a CAN node
recognizes an error condition during a data or remote frame
and issues an active error flag.
• A partial frame occurs when the oscilloscope detects any
error condition during a frame that is not followed by an
active error flag. Partial frames are not counted.
Counters
• The FRAMES counter gives the total number of completed
remote, data, overload, and active error frames.
• The OVLD counter gives the total number of completed
overload frames and their percentage of the total number of
frames.
• The ERR counter gives the total number of completed active
error frames and their percentage of the total number of
frames.
• The UTIL (bus load) indicator measures the percentage of
time the bus is active. The calculation is done on 330 ms
periods, approximately every 400 ms.
Example: If a data frame contains an active error flag, both the
FRAMES counter and the ERR counter will be incremented. If a
data frame contains an error that is not an active error it is
considered a partial frame and no counters are incremented.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSO6012A and is the answer not in the manual?

Agilent Technologies MSO6012A Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSO6012A
CategoryTest Equipment
LanguageEnglish

Related product manuals