Making Measurements 5
6000 Series Oscilloscope User’s Guide 207
Multiply
When you select 1*2, channel 1 and channel 2 voltage values
are multiplied point by point, and the result is displayed.
1 * 2 is useful for seeing power relationships when one of
the channels is proportional to the current.
1 Press the Math key, press the 1*2 softkey, then press the
Scale or Offset sofkeys if you want to change the scaling or
offset for the multiply function.
• Scale — lets you set your own vertical scale factors for
multiply expressed as V
2
/div (Volts-squared/division),
A
2
/div (Amps-squared/division), or W/div
(Watts/division or Volt-Amps/division). Units are set in
the channel Probe menu. Press the Scale softkey, then
turn the Entry knob to rescale 1 * 2.
• Offset — lets you set your own offset for the multiply
math function. The offset value is in V
2
(Volts-squared),
A
2
(Amps-squared), or W (Watts) and is represented by
the center horizontal grid line of the display. Press the
Offset softkey, then turn the Entry knob to change the
offset for 1 * 2.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSO6012A and is the answer not in the manual?

Agilent Technologies MSO6012A Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSO6012A
CategoryTest Equipment
LanguageEnglish

Related product manuals