Making Measurements 5
6000 Series Oscilloscope User’s Guide 213
Integrate
dt (integrate) calculates the integral of the selected source.
You can use integrate to calculate the energy of a pulse in
volt-seconds or measure the area under a waveform.
dt plots the integral of the source using the “Trapezoidal
Rule”. The equation is:
Where
I = integrated waveform
t = point-to-point time difference
y = channel 1, 2, or function 1 + 2, 1 2, and 1 * 2 data
points
c
o
= arbitrary constant
i = data point index
In Delayed sweep horizontal mode, the dt function does not
display in the delayed portion of the display.
1 Press the Math key, press the dt softkey, then press the
Source, Scale, or Offset sofkeys if you want to change the
source, scaling, or offset for the integrate function.
Source — selects the source for dt. The source can be
any analog channel, or math functions 1 + 2, 1 2,
and 1 * 2.
Scale — lets you set your own vertical scale factors for
dt expressed in unit-seconds/division, where units can
be V (Volts), A (Amps), or W (Watts). Units are set in
the channel Probe menu. Press the Scale softkey, then
turn the Entry knob to rescale dt.
n
I
n
= c
o
+ t y
i
i=0

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies MSO6012A and is the answer not in the manual?

Agilent Technologies MSO6012A Specifications

General IconGeneral
BrandAgilent Technologies
ModelMSO6012A
CategoryTest Equipment
LanguageEnglish

Related product manuals