358 6000 Series Oscilloscope User’s Guide
9 Characteristics and Specifications
Trigger system
Sources DSO6xx4L: Ch 1, 2, 3, 4, line, ext and D0 - D15 for
MSO enabled DSO
Modes Auto, Normal, Single
Holdoff time range ~60 ns to 10 seconds
Trigger jitter 15 psec rms
Selections Edge, pulse width, pattern, TV, duration,
sequence, CAN, LIN, USB, I2C, SPI, Nth edge
burst
Scope channel triggering
Range (internal) ±6 div from center screen
Sensitivity* < 10 mV/div: greater of 1 div or 5 mV
≥ 10mV/div: 0.6 div
Coupling AC (~10 Hz), DC, noise reject, HF reject and LF
reject (~ 50 kHz)
* Denotes warranted specifications, all others are typical. Specifications are valid
after a 30-minute warm-up period and ±10 °C from firmware calibration
temperature.
Logic (D15 - D0) channel triggering (with MSO option)
Threshold range (user defined) ±8.0 V in 10 mV increments
Threshold accuracy* ±(100 mV + 3% of threshold setting)
Predefined thresholds TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V
* Denotes warranted specifications, all others are typical. Specifications are valid
after a 30-minute warm-up period and ±10 °C from firmware calibration
temperature.
External (EXT) triggering
Input resistance 1.015 kΩ ± 5% (DSO6014L)
2.14 kΩ ± 5% (DSO6054L/6104L)
Maximum input ±15 V
Range ±5 V
Sensitivity DC to 100 MHz: 500 mV (DSO6014L)
DC to 500 MHz: 500 mV (DSO6054L/6104L)
Coupling AC (~ 3.5 Hz), DC, noise reject, HF reject and LF
reject (~ 50 kHz)
Probe ID Auto probe sense (DSO6014L)
Auto probe sense and AutoProbe interface
(DSO6054L/6104L)