Chapter 4 CATALYS® System Procedure
Page 174 of 314 Part Number: 0160-6411, Rev A, 02/2019
Software Release: cOS 5.00
Table 4.6 User-adjustable Parameters for Primary and Sideport Cataract Incisions
Feature Default
a
a.Parameters do not have default values; the user must select each parameter.
Range Increment Step Size Units
Number of Primary Incisions, n
Number of Sideport Incisions, n
N/A
N/A
1-2
0-5
1
1
1
1
N/A
N/A
Axis N/A 0-360 1 5 °
Limbus Offset N/A 0.0-5.0 0.1 0.1 mm
Width N/A 0.2-6.5 0.1 0.1 mm
Length N/A 0.5-3.0 0.1 0.1 mm
Uncut Region Central
Anterior, Central,
Posterior, None
N/A N/A N/A
Uncut Length
Anterior
Posterior
20% (100 µm)
20% (100 µm)
20-50% (100-250 µm)
20-50% (100-250 µm)
1% (1 µm)
5% (25 µm)
5% (25 µm)
%, µm
Uncut Central Length 25 25-1000 1 25 µm
Plane Depth
Anterior
Posterior
30% (150 µm)
70% (350 µm)
25-75% (125-375 µm)
25-75% (125-375 µm)
1% (1 µm)
1% (1 µm)
5% (25 µm)
50% (250 µm)
% (µm)
Side Cut Angle
Anterior
Posterior
90
45
30-150
30-150
1
1
5
5
°
°
Line Density
Anterior
Central
Posterior
10
1
4
1-10
1-10
1-10
1
1
1
1
1
1
N/A
N/A
N/A
Line Distance
b
Anterior
Posterior
b.Note that the central line distance is not user‐adjustable. (It is the difference between the sum of anterior
and posterior line distances.)
30% (150 µm)
30% (150 µm)
0-75% (0-375 µm)
0-75% (0-375 µm)
1% (1 µm)
1% (1 µm)
5% (25 µm)
5% (25 µm)
% (µm)
Spot Spacing
Horizontal
Vertical
5
10
3-50
5-50
1
1
5
5
µm
µm
Pulse Energy 5 3-10 0.1 0.5 µJ