Automation
Cartridge sipper system
72 Edition 09/2014 SPECORD S600 accessories
On the tab Accessories the following configurations are carried out:
Option /
button
Description
Pump time [s]
The pumping time is the time required to transport the
sample optimally into the cell. To optimize the flushing
effect approx. 2/3 of the sample should flow through the
cell and 1/3 of the sample should be upstream of the
cell.
Adjustment range: 0 ... 300 s
Wash before
measurement
Stipulation of a flushing step before the sample
measurement.
The flushing time is 1.5 x pumping time.
The times are rounded to tenths of a second after
entering the figure.
Wash after
measurement
Stipulation of an additional flushing step after the sample
measurement.
[Pump]
Switching on the pump to check aspiration or to
determine the optimum pumping time.
The label of the button changes dependent on the
switching state of the pump.
Carrying out the sample measurement
Reference measurement
Dependent on the stipulation in the measuring parameters carry out a reference
measurement.
Start the reference measurement with or the menu command
Measurement / Reference Measurement. Proceed as described under
sample measurement.
Sample measurement without additional flushing steps
1. Move the measuring sample to the intake hose.
2. Start the measurement with or the menu command Measurement /
Measurement. The sample is aspired and the measurement starts.
Sample measurement with the option "Wash prior to measurement"
1. Press .
2. A window opens where you are prompted to move the flushing container to
the intake hose.
3. Move a container with flushing liquid to the intake hose.
4. Start the flushing process with [OK].
5. If the flushing process is complete the window "Prepare measuring sample!"
appears.
6. Move your sample to the intake hose. Start the measurement with [[OK].
The sample is aspired and the measurement starts.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analytik Jena SPECORD S600 and is the answer not in the manual?

Analytik Jena SPECORD S600 Specifications

General IconGeneral
BrandAnalytik Jena
ModelSPECORD S600
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals