Automation
Cartridge sipper system
SPECORD S600 accessories Edition 09/2014 73
Sample measurement with the option "Wash after measurement"
1. Move the sample to the intake hose.
2. Start the measurement with .
3. The sample is aspired and the measurement starts.
4. If the measuring process is complete the window "Prepare flushing sample!"
appears.
5. Move a container with flushing liquid to the intake hose
6. Start the flushing process with [OK].
Care and maintenance
Your maintenance tasks for the cartridge sipper system are limited to replacing
the pump hoses. Pump hoses from different materials or with different inside
hose diameters are available from our customer service.
Observe the following maintenance instructions:
Avoid contamination at the sample holder. Wipe spilled samples or
reagents immediately with an absorbent cloth or piece of paper. Remove
stubborn contamination with a soft cloth wetted with a commercial
dishwashing detergent.
After completing the work fill the flow cell with distilled water. If the cell
dries out completely any sample residue might get stuck in the flow cell.
After completing the work relieve the pump hose by releasing the hose
cartridge from the pump. This ensures that the pump hoses remain
elastic for longer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analytik Jena SPECORD S600 and is the answer not in the manual?

Analytik Jena SPECORD S600 Specifications

General IconGeneral
BrandAnalytik Jena
ModelSPECORD S600
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals