CPS100 SERIES OPERATION MANUAL
SECTION 6:
Front Panel Operation
Entire Contents Copyright 2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CPS100 Series Power Source Operation Manual Page 132 of 231
IEC 4-13 Meister Curve Test
From the top level IEC 4-13 menu, press the
Page Select soft key to access the second
screen of the IEC 4-13 tests. Next, press the
Meister Curve soft key to display the Meister
Curve test screen.
Available soft keys in this screen are Trigger
and Edit. Press the Edit soft key to set the
required test selections for this test.
There are two standard product classes for
the IEC 61000-4-13 Meister Curve test. They
are Class 2 and Class 3. (Standard modes)
Test levels and durations for these product
classes are defined by the IEC 61000-4-13
standard. There is also a User class, which
allows parameters to be changed. (User
mode).
To select the appropriate product class for
the unit under test, move the cursor to the
“Level =” field and press [Enter]. A drop
down list with product class selections is
displayed.