6.7.2.2 Average Setting
The Average soft key changes the number of
times a reading is averaged before the
corresponding measurement parameter
display is updated. Available average values
are 1, 2, 4, 8, 16 and 32.
The higher the number readings averaged,
the slower the measurement update rate will
be so we recommend you choose the lowest
acceptable number.
Make a selection by pressing the Average
soft key and selecting an averaging value
from the list, then press ENTER to confirm.
6.7.2.3 Is Delay & Is Interval Setting
The Is Delay and Is Interval settings are used
to determine the inrush current of a unit
under test.
Is delay defines the delay time between the
beginning of the output voltage transition to
the start of the measurement of I surge.
Is Interval defines the measurement time.
For example, for a 50Hz application, setting
the Is delay to 10 msec will cause the I surge
current measurement to start half a cycle
after the voltage is applied.
Available setting values are:
Is Delay: 0.0 ~ 999.9 msec
Is Interval: 0.0 ~ 999.9 msec