EasyManua.ls Logo

biodex ATOMLAB 500 - Page 172

biodex ATOMLAB 500
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
!
172 ATOMLAB 500 DOSE CALIBRATOR
To Perform A Manual Decay Test:
1. From the Linearity Testing Menu, touch <Manual Decay Test>. The Manual Decay Test
screen is now displayed.
• Touch <Back> to return to the Linearity Testing screen.
• Touch <Select Detector> to select a different detector if multiple detectors are selected.
• Touch <Select Decay Isotope> to select the isotope to be used in the decay test.
• Touch <Select Setup Options> to select setup options.
• Touch <Zero Background> to zero the background reading if necessary.
2. Touch <Start Test> to begin the manual decay test. The Manual Decay Test screen is displayed.
3. Place the source to be used into a dipper and then place the dipper with the source into
the chamber.
4. Once the reading has stabilized, touch <Record Reading>. The reading is recorded and the
next line for the manual decay test is displayed. The system will continue to display a live
reading until <Record Reading> is selected.
5. Continue taking readings at the appropriate time intervals by touching <Record Reading>.
NOTE: If you need to use the calibrator for counting doses, touch <Measure> to advance to
the Measure screen. Touch <Return to Manual Decay Test> to go back to the manual
decay test.
6. Touch <End Test> to end the test and advance to the Normalized Reading screen.
Figure 6.43. The Manual Decay Test screen with Test Ended message and data displayed

Table of Contents

Other manuals for biodex ATOMLAB 500

Related product manuals