P1 Patient Monitor User’s Manual
14-5
2. Install another end of the sampling tube to patient.
CAUTION:
Pay attention to the water level of water trap. If the highest water level reaches,
please replace the water trap in time to prevent the module from entering by
water.
Please keep the sampling tube clean, and prevent the tube from clogging by
dust.
When using a sidestream CO
2
module, the exhaust holes on the module must be
connected to the exhaust gas treatment system.
NOTE
:
Inserting the CO
2
module into slot, the monitor automatically starts the
sampling pump. Removing the CO
2
module can turn the sample pump off.
Water trap and sampling tubes are disposable, please use accessories provided
or designated by manufacturer.
14.4.3. Measure Using Microflow CO
2
module
1. Attaching the catheter, airway adapter or sampling tube
Connect the catheter, airway adapter or sampling tube to the CO
2
plug-in module as
required. Take Nomoline ISA CO
2
module as an example, as shown in the following
figure:
Sampling
tube port