P1 Patient Monitor User’s Manual
17-9
CAUTION:
When ECG lead is set 3 lead, ECG report cannot be printed.
17.3.5.2. Setting the Realtime Reports
To set realtime reports, follow this procedure:
1. SelectMain Menuquick keyfromReportcolumn to selectReport
Setup.
2. Select Realtime Report.
3. Set the desired items. The following table only lists some of the options.
Menu item Function Description
Select
Waveform
Select the desired waveform
to printout
Current Waveformsprints the
realtime report for current waveforms.
Selected Waveformsprints the
realtime report for the selected
waveforms.
17.3.5.3. Setting Tabular Trends Reports
To set tabular trends reports, follow this procedure:
1. SelectMain Menuquick keyfromReportcolumn to selectReport
Setup.
2. Select Tabular Trends Report.
3. Set the desired items. The following table only lists some of the options.
Menu item Function Description
Period Select the specified period
during which a tabular
trends report will be printed.
Auto】:print one page of tabular trends
report according to the selected resolution.