P1 Patient Monitor User’s Manual
3-11
The monitor behaves as follows after entering the standby mode:
Stops all parameter measurements.
Disables all the alarms and prompt messages, except for the battery low
alarm.
Turns screen brightness to the dimmest after entering the standby mode
for 30 seconds.
WARNING:
Pay attention to the potential risk of placing the monitor to standby. In the
standby mode, the monitor stops all parameters’ measurements and disables
all the alarm indications, except for the battery low alarm.
3.6.3.2. Exit the Standby Mode
To exit the standby mode, choose any of the following ways:
Select
【
Resume Monitor
】
to exit the standby mode and resume monitoring the
current patient.
Select
【Discharge Patient】to discharge the current patient.
If the monitor automatically enters the standby mode after a patient is discharged,
choose any of the following ways to exit the standby mode:
Select【Patient Management】 to exit the standby mode and admit a new
patient.
Select【Monitor】to enter the patient information for preparing to admit a new
patient.
3.6.4. Demo Mode
In Demo mode, the monitor can demonstrate its major functions when patient or
patient simulator is not connected. The Demo mode is password protected
Choose the following way to enter the Demo Mode:
1. Select【Main Menu】from【System】column to select【Demo】.
2. Input the passwordselect【OK】.
Shut down and restart to exit the Demo Mode.