P1 Patient Monitor User’s Manual
3-12
WARNING
The demonstration function is mainly used to display machine performance
and to train users. In the actual clinical use, it is forbidden to use the
demonstration function, so as to prevent the medical staff from mistakenly
thinking that the monitor displays the waveform and parameters of the
monitored patient, thus affecting the patient’s monitoring and delaying the
diagnosis and treatment.
3.6.5. Night Mode
Night mode is a special clinical monitoring mode. Under the night mode, the alarm
volume, QRS volume and screen brightness turn to be lowest automatically.
To avoid
disturbing the patient, night mode may be used.
3.6.5.1. Entering Night Mode
The steps of enteringNight Modeas following:
1. SelectNight Modequick key or selectMain Menuquick keyfrom
Displaycolumn to selectNight Modebutton.
2. In the pop-up menu, set thenight mode.
3. SelectEnter Night Mode.
The night mode settings are as follows by default:
Brightness: 1
Alarm Volume: 2
QRS Volume: 1
Touch Tone: Off