42 (167) BRUKER Installation Manual Version 001
Software Commands
Standard Commands
Table 7.1. Standard Commands
Instruction: AP (AutoPrep)
Format: AP<CR>
Description: AutoPrep and barcode reader present test.
Reply: <CR><LF>
______________________________________________________________
Instruction: EJ (EJect)
Format: EJ<CR>
Description: Get sample from shim system and insert it in the magazine.
Reply: <CR><LF>
______________________________________________________________
Instruction: HO (HOme)
Format: HO<CR>
Description: The sample changer arm moves to its „HOME“ position over
the magazine. If a sample is in the pincher, it is placed in the
magazine when the given position is free
Reply: <CR><LF>
______________________________________________________________
Instruction: IJ (InJect)
Format: IJ XXX<CR>
Description: Get sample from position XXX from the magazine and insert it
in the shim system.
Reply: <CR><LF>
______________________________________________________________
Instruction: NM (total Number of Magazine positions)
Format: NM<CR>
Description: Report the number of available magazine positions.
Reply: N<number><CR><LF>
______________________________________________________________
Instruction: RP (Report Position)
Format: RP<CR>
Description: Read actual magazine position.
Reply: P<number><CR><LF>