B-ACS 60/120 Command Implementation
Installation Manual Version 001 BRUKER 43 (167)
______________________________________________________________
Instruction: SE (SEcurity)
Format: SE<CR>
Description: Security Test. Start a sample lift lasting 30 sec's to ensure that
there is no sample in the magnet. If a sample is found, the
sample lift is kept on and the message S1 is sent back. If no
sample is found, the message given is: "S0".
Reply: S0<CR><LF> or S1<CR><LF>
______________________________________________________________
Instruction: SP (Sample Present in magazine)
Format: SP XXX<CR>
Description: Report sample status of magazine position XXX. The mes-
sage "S1" means sample found, "S0" means no sample found.
Reply: S0<CR><LF> or S1<CR><LF>
______________________________________________________________
Instruction: VS (Version of Software)
Format: VS<CR>
Description: Report Firmware version. Format YYMMDD, which means
YEAR MONTH DAY. Example 991013.
Reply: YYMMDD<CR><LF>
______________________________________________________________
Instruction: ZY (Zero Yell)
Format: ZY<CR>
Description: Repeat the answer of the last command.
Reply: Last reply followed by <CR><LF>
______________________________________________________________
Instruction: ESC. (ESCape key + ".")
Format: ESC.<CR>
Description: Switch to the Diagnostic mode.
Reply: SAMPLECHANGER DIAGNOSTIC MODE!
Table 7.1. Standard Commands