B-ACS 60/120 Command Implementation
Installation Manual Version 001 BRUKER 47 (167)
Instruction: A1 (Arm to magazine position 1)
Format: A1<CR>
Description: Move to inner magazine position. If the pincher was in the
lower position, the arm will move up before moving to the inner
magazine position.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: A2 (Arm to position 2)
Format: A2<CR>
Description: Move to outer magazine position. If the pincher was in the
lower position, the arm will move up before moving to the outer
magazine position.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: BT (Barcode Test)
Format: BT<CR>
Description: Barcode reader test. Start the bar code reader test. the mes-
sage "TEST CHANNEL A" shows up. Use the front panel keys
"<" and ">" to switch from channel A to channel B and back.
The active channel is always displayed. Pressing the CONT
button stops the test.
Reply: EXECUTED!<CR>
______________________________________________________________
Instruction: CA (Code wheel Adjust)
Format: CA<CR>
Description: Code wheel adjust. The code wheel will be read and shown
continuously on the LC display. Pressing the CONT button
stops the test.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: DT (Display Test)
Format: DT<CR>
Description: Display test. Calls the display test program with the character
pattern check routine. Pressing the CONT button stops the
test.
Reply: EXECUTED!<CR><LF>
Table 7.4. Diagnostic Mode Commands