50 (167) BRUKER Installation Manual Version 001
Software Commands
Instruction: PC (Pincher Close)
Format: PC<CR>
Description: Close pincher.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: PL (Pincher Loop)
Format: PL<CR>
Description: Test loop for pincher movement. Loop count will be displayed
on the terminal. Pressing the CONT button when the red LED
is on stops the test.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: PO (Pincher Open)
Format: PO<CR>
Description: Open pincher.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: PT (Pressure Test)
Format: PT<CR>
Description: Pressure test. This command starts a loop program which
checks continuously the condition of the pressure switch and
indicates it on the LC display. If the air pressure exceeds the
threshold value, the message displayed will be:
"SUFFICIENT AIR PRESSURE". If the pressure falls below
the threshold, the message given on the display is:
"INSUFFICIENT AIR PRESSURE". Pressing the CONT button
stops the test.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: TL (Test Loop)
Format: TL<CR>
Description: Complete sample changing test loop. The magazine moves on
until a sample has been found and then a sample change is
executed. The number of executed sample changes will be
displayed on the terminal. Pressing the CONT button when the
red LED is on stops the test.
Reply: EXECUTED!<CR><LF>
Table 7.4. Diagnostic Mode Commands