B-ACS 60/120 Command Implementation
Installation Manual Version 001 BRUKER 49 (167)
______________________________________________________________
Instruction: MF (Magazine one step Forwards)
Format: MF<CR>
Description: Move magazine one position forward.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: MP (Move to Position)
Format: MP XXX<CR>
Description: Move magazine to position XXX.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: OH (Optics on Horizontal cylinder)
Format: OH<CR>
Description: Test for light barriers for horizontal positioning. Pressing the
CONT button stops the test. The message given on the dis-
play during the optic test is: REFLECTION OK 1 or
NO REFLECTION 0.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
Instruction: OM (Optics at Magazine)
Format: OM<CR>
Description: Test for light barriers at magazine. Pressing the CONT button
stops the test. The message given on the display during the
optic test is: SAMPLE PRESENT 1 or SAMPLE MISSING 0.
Reply: EXECUTED!<CR><LF>
______________________________________________________________
OS (Optic Shim)
OS<CR>
Test for light barrier at the shim system. Pressing the CONT
button stops the test. The message given on the display during
the optic test is: SAMPLE PRESENT 1 or
SAMPLE MISSING 0.
EXECUTED!<CR><LF>
______________________________________________________________
Table 7.4. Diagnostic Mode Commands