125
Z31900_00_01
Safety Notices
NOTICE
Material damage hazard due to impacting the magnet.
Impacting the magnet may result in a quench.
f Mount the device carefully on the magnet.
f Avoid banging the magnet during installation and operation, e.g. when replacing
the sample cassette.
NOTICE
Material damage hazard from unsafe assembly practices!
Material damage may result from falling from the ladder during assembly, retrofitting,
or dismantling. This may result in personal injury or equipment damage.
f Use a platform with railings instead of a ladder to reach the assembly area.
f Where appropriate clothing, e.g. slip-free shoes.
NOTICE
Material damage hazard from illicit sample movement!
Material damage may result from illicit sample movement during installation and
maintenance.
f Use only dummy samples during installation and maintenance.
WARNING
Accident hazard from suffocation!
A break in the pneumatic hose may result in the uncontrolled exit of nitrogen into the
laboratory.
f An oxygen warning device should be present in the laboratory if the device is
operated with nitrogen.
f Note that leakage from the main supply line cannot be stopped by the device!